Particle characterisation
Descriptions of our particle characterisation instrumentation, including specifications and typical applications.
Malvern Mastersizer 3000 PSA
Our Mastersizer system allows particle size distribution to be measured via laser diffraction. A laser is passed through a dispersed particulate sample and the intensity of light scattered at different angles recorded. This diffraction pattern is correlated with modelled size distributions.
Our current capabilities are for wet dispersion only, in either water or IPA. The system has a measurement range of 0.01μm - 2mm when using wet dispersion.
Applications
Particle size distribution has a direct effect on various material properties, such as
- reactivity or dissolution rate
- stability in suspension
- efficiency of delivery
- viscosity
- flowability
Specification
Particle size range | 0.01 μm to 2mm (for wet dispersion) |
---|---|
Accuracy | Better than 1% |
Repeatability | Better than 0.5% variation |
Reproducibility | Better than 1% variation |
Detection systems |
Red light - forward, side and back scattering Blue light - wide angle forward and back scattering |
Dispersion | Wet (water or IPA) |
Measurement principle | Mie and Fraunhofer scattering |
Micromeritics AccuPyc II 1340
The Micromeritics helium gas displacement pycnometer allows the non-destructive measurement of volume and density of powder and solid materials. An inert gas (helium) is used as a displacement medium. Since the sample is sealed a compartment of known volume, when the purge gas is admitted and then expanded into a second internal volume, the pressure before and after expansion can be measured and used to determine sample volume.
Absolute (or true) density, ie the density obtained where the volume measured excludes pores and voids between particles, is determined using helium pycnometry.
Applications
- Density determination
- Volume determination
Specifications
Precision | +/- 0.01% (typical) +/- 0.02% (guaranteed) |
---|---|
Accuracy | Within 0.03% reading, plus 0.03% of sample capacity |
Gas | Research grade helium |
Sample cell size | 1cm3 or 0.1cm3 available |
Temperature | Not controlled |